Catalog & Support

  • Instrumentation & Services

    Find Products, Service Centers, Dealerships, Manuals, and more.

  • Business Services

Companies & Community

  • Company
  • Industry Forums
  • T&M Job Bank (Openings)

Classifieds

  • Products & Services
  • Assets for Share/Rent
  • Trade-Ins & Deals
  • Donations & Recycle
  • Books

    Offered

    Wanted

Support

  • Product Documentation
  • Training
  • Techniques & Solutions
  • Product & Service Advisor

Resources

Social

  • Forums
  • Global Event Calendar
  • Frequently Asked Questions

____ This section requires a Corporate membership ____

Products & Services Wanted

  • Products & Related Services
  • Trade-Ins
  • Donations & Recycle
  • Open Solicitations - Products
  • Open Solicitations - Services

Mergers & Acquisitions

  • Offered
  • Wanted

Collaboration

  • Collaboration Offered
  • Collaboration Wanted
  • Sales Channels Offered
  • Sales Channels Wanted
  • T&M Job Bank (Resumes)
Stacks Image 161208

Miscellaneous Industries - Product Spotlight

Guildline Instruments
8400B
Salinometer

The 8400B “Autosal” Laboratory Salinometer is the “Industry Standard” for salinity measurements. A very stable temperature control bath and heat exchanger maintain the sample at a precisely defined temperature during analysis, avoiding the need for temperature compensation. Accuracy is better than 0.002 Equivalent Practical Salinity Units (PSU).

PRODUCT FEATURES

  • Accuracy <0.002 Salinity
  • Range 2 to 42 in Salinity
  • Improved Temperature Stability
  • No Temperature Compensation Required
  • Small Samples & Rapid Measurements
  • Data Logging Standard
  • Improved Maintainability

Measurement Range:0.0001:1.15 Conductivity Ratio, 0.004 to 76mS/cm, 2 to 42 Equivalent Practical Salinity Units (PSU)

Accuracy:< ± 0.0001 Conductivity Ratio, @ same set point temperature as standardization and within -2 0C and +4 0C of ambient. By calculation & substitution in the Bennett equation or the UNESCO tables, < ± 0.002 Equivalent PSU.

Short Term Stability:< ± 0.00005 for 24 hours without restandardization

Maximum Resolution: < 0.00001 Conductivity Ratio< 0.0002 mS/cm @ 15 0C and 35 PSU < 0.0002 Equivalent PSU

Sample Volume: Maximum required – 100 milliliters (starting from fresh water in the cell), including flushing volume. About 50 milliliters for 3 Equivalent PSU difference in samples.

Scale Suppression: Linear scale of conductivity ratio having 22 steps from 0 to 2.2, where 2.0 corresponds to seawater of 35 Equivalent PSU. Maximum reading is 2.29999, corresponding to approximately 42 Equivalent PSU.

Bath Temperature: Selectable from 18 0C to 33 0C in 3 0C steps, accuracy ±0.02 0C, stability ± 0.0010C per day. Selected temperature should be within (ambient +4) 0C and (ambient -2) 0C.

Outputs: TTL compatible BCD outputs of numerical display readings and thumbwheel switch bottle logger.

Maintenance:Cell easily cleanable by bottle brush. Can be removed for cleaning and replaced without change in calibration.

Water Bath Volume: 4.4 U.S. gallons (16.8 liters).

Power Required: 115 or 230 volts ± 10%, 50/60 Hz, 400 watts maximum

Temperature: Operating: 16 0C to 37 0C Storage: -40 0C to 70 0C

Dimensions: H 678 mm (26.7 in); W 533 mm (21 in); D 559 mm (22 in)

Weight:Bath Empty: 50 kg (110 lbs)

— no reviews —

Test Interest LLC
5021 Verdugo Way, Suite 105 #106
Camarillo
California
93012
United States
Listings
New
0
0
Used
0
0
Rent-lease
0
0
Auction
0
0
Trade-Ins
0
0
Donations
0
0
Authorized Dealers
0
0

All Registered Trademarks, brand names and logos are respected and their ownership acknowledged. They are used on this website for information and reference purposes only. Test Interest doesn't assume any other implication or representation.

TI-IND-MIS-3